單光子雪崩二極體應用於 X 光與高能質子偵測

X-ray and Energetic Proton Detections Using Single-Photon Avalanche Diodes

檔案

全文將於 2028年12月1日 起公開。

日期

2023

類型

語言

中文

摘要

CMOS 單光子雪崩二極體具有低暗計數、高靈敏度、高時間解析度及低製造成本的優點。在本研究中,針對兩種輻射源進行偵測,其一以國家同步輻射研究中心-臺灣光子源的脈衝式X光作為輻射源,使用碘化銫作為閃爍體蒸鍍在SPAD表面,透過SPAD吸收碘化銫的放光,間接偵測X光。成功量測臺灣光子源的時間響應,可以清楚分辨multiple bunches每間隔2 ns的脈衝,推論鍍製碘化銫的SPAD之時間解析度至少在2 ns以下。在目前的實驗架設中,每秒最低只需10個X光光子,即可分辨由X光引起的光計數,並透過換算得到0.075 nGy/s的最低可測劑量,優於目前所有文獻的X光偵測器。
其二以清華大學加速器館的500 kV離子佈植實驗站產生的質子束作為輻射源,採取質子直接打在SPAD所產生的崩潰事件進行量測,在質子束能量300 keV以及電流100 nA時成功量測到由質子引起的計數,SPAD在超額偏壓2 V的質子偵測機率約2.97 %,為首次以SPAD直接偵測質子的量化結果,期望SPAD 在輻射相關領域也能成為新興的感測器。
Single-photon avalanche diodes (SPAD) in CMOS technology have low noise, high sensitivity, excellent timing resolution and low fabrication cost. In this study, two radiation sources were detected. One source was the pulse X-ray from the Taiwan photon source in the National Synchrotron Radiation Research Center. Using cesium iodide (CsI) as the scintillator coated on the surface of the SPAD, the fluorescence from CsI is absorbed by the SPAD, indirectly detecting X-ray. The successful measurement of the time response of the Taiwan photon source revealed the ability to clearly distinguish pulses occurring at intervals of 2 ns, indicating that the time resolution of the SPAD was at least below 2 ns. With my experimental setup, the minimum of ten X-ray photons per second is required to distinguish the photon counts induced by X-ray. Through conversion, the minimum dose rate detection limit is approximately 0.075 nGy/s, surpassing all X-ray detectors.
The second part of the study used proton beams as the radiation source, generated by the 500 kV ion implanter at the National Tsing Hua University. Measuring the breakdown events produced by the SPAD through proton hit directly. Successful detection of the proton induced events was achieved at a proton beam energy of 300 keV and a current of 100 nA. The proton detection probability of the SPAD at an excess bias of 2 V is approximately 2.97 %. The first quantified results of directly detecting protons using SPAD. Hope the SPAD can also become a new option of radiation-related sensors.

所屬典藏

電子研究所
顯示完整詮釋資料 統計資料

引用此條目

呂孟軒 (2023). 單光子雪崩二極體應用於 X 光與高能質子偵測. https://thesis.lib.nycu.edu.tw/handle/20.500.14371/990

BibTeX RIS

引用文字依 APA 第 7 版格式自動產生,使用前請再核對。

分享

永久網址:https://thesis.lib.nycu.edu.tw/handle/20.500.14371/990